Determination of iron in blood serum using flow injection with luminol chemiluminescence detection.
Authors:A Waseem  M Yaqoob  A Nabi
Affiliation:Department of Chemistry, University of Baluchistan, Quetta, Pakistan. waseemq2000@yahoo.com
Abstract:A simple and rapid fl ow injection method is reported for the determination of iron in blood serum after acid digestion with HNO3 and HClO4, based on luminol CL detection in the absence of added oxidant. The detection limit (3 s) was 1.0 nmol/L with a sample throughput of 120/h. The calibration graph was linear over the range 0.001-1.0 micromol/L (r2 = 0.9974), with relative standard deviations (RSD) (n = 4) in the range 3.2-5%. The effect of interfering cations (Ca(II), Mg(II), Cu(II), Cd(II), Pb(II), Mn(II), Zn(II), Ni(II), Co(II) and Fe(III)) and anions (Cl-, SO4(2-), HCO3-, NO3-, NO2-) were studied using a luminol CL system for Fe(II) determination. The method was applied to normal blood serum and the results (1.32 +/- 0.08-1.74 +/- 0.05 mg/L) were compared with those from a spectrophotometric reference method (1.34 +/- 0.06-1.80 +/- 0.10 mg/L), which agree fairly well with the overall reference range in blood.
Keywords:flow injection analysis  chemiluminescence  iron  blood serum