Abstract:The procedures and equipment needed to measure filtration efficiency by means of fluorescent aerosols are described. The filtration efficiency of individual lots of spun-glass air filtration medium or of entire air filtration systems employing such media was determined. Data relating to the comparative evaluation of spun-glass filter media by means of the fluorometric method described, as well as by conventional biological procedures, are presented.