Comparison of calibration methods for respiratory inductive plethysmography in infants |
| |
Authors: | Revow, M. D. England, S. J. Stogryn, H. A. Wilkes, D. L. |
| |
Affiliation: | Institute of Biomedical Engineering, University of Toronto, Ontario, Canada. |
| |
Abstract: | In infants under the age of 6 mo respiratory inductive plethysmograph (RIP)-calculated tidal volumes (VT) were compared with simultaneously measured volumes using a pneumotachograph (PNT) to 1) assess whether using multiple points (MP) along the inspiratory profile of a breath is superior to using only VT when calculating volume-motion (VM) coefficients, 2) verify the assumption of independent contributions of the abdomen and rib cage to VT, which was accomplished by extending the normal RIP model to include a term representing interaction between these two compartments, and 3) investigate whether VM coefficients are sleep-state dependent. Neither use of multiple points nor inclusion of the interacting term improved the performance of the RIP over that observed using a simple two-compartment model with VT measurements. However, VM coefficients obtained during quiet sleep (QS) were not reliable when used during rapid-eye-movement (REM) sleep, suggesting that coefficients obtained during one sleep state may not be applicable to another state where there is a substantial change in the relative abdominal/rib cage contributions to VT. |
| |