Immunodetection of pentamer and modified C-reactive protein using surface plasmon resonance biosensing
Authors:Hu W P  Hsu H-Y  Chiou A  Tseng K Y  Lin H-Y  Chang G L  Chen S-J
Institution:Institute of Biomedical Engineering, National Cheng Kung University, Tainan, Taiwan.
Abstract:In clinical practices, the examination of pentamer C-reactive protein (pCRP) is commonly used as a prognostic indicator of the risk of a patient developing cardiovascular disease (CVD). Structural modification of pCRP produces a modified CRP (mCRP) which exhibits different biological activities in the body. In recent years, mCRP has come to be regarded as a more powerful inducer than pCRP, and hence mCRP measurement has emerged as an important indicator for assessing the risk of developing CVD. The surface plasmon resonance (SPR) biosensing technique can be employed to increase the detection accuracy and real-time response when sensing pCRP or mCRP. In this study, three monoclonal antibodies (Mabs), C8, 8D8, and 9C9, are immobilized on a protein G layer for subsequent CRP detection. The experimental results reveal that the Mab C8 reacts with both pCRP and mCRP, the Mab 8D8 with pCRP, and the Mab 9C9 with mCRP. No false signals caused by non-specific binding are observed. When detecting pCRP using Mab C8, the SPR bioassay provides sufficient sensitivity to evaluate whether or not a patient is at risk of developing CVD. SPR biosensing provides a viable and accurate approach for the real-time evaluation of pCRP and mCRP levels, and is therefore of considerable benefit in clinical examinations of CPR.