首页 | 本学科首页   官方微博 | 高级检索  
     


Column-switching high-performance liquid chromatographic method for the determination of a thymidylate synthase inhibitor, LY231514, an investigational agent for the treatment of solid tumors, in human plasma
Authors:Cristi L. Hamilton  Jeffrey A. Kirkwood
Abstract:A reversed-phase, column-switching high-performance liquid chromatographic (HPLC) method is described for the determination of a new thymidylate synthase inhibitor in human plasma. The compound and an internal standard are extracted from plasma using a Certify II solid-phase cartridge. Extracts are evaporated to dryness and the residue is reconstituted with mobile phase buffer. The analytes are separated from polar interferences and buffer salts originating from the elution step on a 4-mm YMC Basic pre-column. The fraction containing the analytes is further separated on a 25-cm YMC Basic column. The analytes are detected by their absorbance at 250 nm. The limit of quantitation is 10 ng/ml. The method is linear from 10 ng/ml to 80 μg/ml using three standard curve ranges. Validation studies for all three ranges show the method to be reproducible. The method has been successfully used to support pharmacokinetic studies.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号