首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Dual chamber pacing mode in an atrial antitachycardia pacing device without a ventricular lead – A necessary evil
Authors:Amit Noheria  Paul A Friedman  Samuel J Asirvatham  Christopher J McLeod
Institution:Division of Cardiovascular Diseases, Mayo Clinic, Rochester, MN, USA
Abstract:We present a case of a single chamber atrial pacemaker implanted for sinus node dysfunction and treatment of macroreentrant atrial tachycardias with atrial antitachycardia pacing. The patient presented with sustained atrial tachycardia above the detection rate, however, the device was unable to detect the tachycardia and did not deliver the programmed therapy. We discuss the nuances of the atrial tachyarrhythmia detection algorithms, and the programming strategies to maximize detection of atrial arrhythmias in a single chamber atrial pacemaker.
Keywords:Pacing  Atrial antitachycardia pacing  Pacing mode  Device algorithm