首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Laccase electrode for the continuous-flow determination of phenolic compounds
Authors:Nabil Zouari  Jean-Loius Romette  Daniel Thomas
Institution:(1) Laboratoire de Technologie Enzymatique U.A. ncompfn523 du CNRS, 60206 Compiègne, France;(2) Present address: Centre de Biotechnologie de Sfax, Sfax, Tunisia
Abstract:Summary Laccase (p-diphenol, O2 oxido-reductase, E.C. 1.10.3.2) from Botrytis cinerea was immobilized in a gelatin support on an O2 sensing electrode. The enzyme was copolymerized with the inert protein using glutaraldehyde (1.25 % w/w) on the hydrophobic selective gas membrane of a pO2 sensor and this was covered with a Nuclepore polycarbonate microporous film (0.03 mgrm). The enzyme electrode was used in a continuous-flow system to measure the concentration of a wide range of phenolic substrates. The measuring time of each sample was about 1.5 min including response and rinsing times. The electrode response was set for hydroquinone up to 0.8 mM with high reproducibility and less than 5 % error.The electrode response for hydroquinone concentration of 0.25 mM was stable with repeated use for at least 800 assays without significant loss of activity.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏