Simultaneous measurement of allantoin, uric acid, xanthine and hypoxanthine in blood by high-performance liquid chromatography |
| |
Authors: | M. Czauderna J. Kowalczyk |
| |
Abstract: | A high-performance liquid chromatographic method for determining catabolism products of nucleic acids and purines, such as oxypurines (i.e. uric acid, xanthine and hypoxanthine) and allantoin in the blood plasma of ruminants was developed. The plasma was deproteinized with 10% trichloroacetic acid. The method enabled determination of oxypurines without derivatization. Allantoin was determined after conversion with 2,4-dinitrophenylhydrazine to a hydrazone (GLX-DNPH). Separation of converted allantoin, uric acid, xanthine and hypoxanthine derivatives was carried out using two reversed-phase C18 columns. The combination of pre-column derivatization and gradient elution with monitoring of the effluent at 205, 254 and 360 nm provides a simple and selective analytical tool for studying oxypurines and allantoin in plasma. The total run time of the HPLC analysis was 60 min. The recovery of the purine derivatives (i.e. oxypurines and allantoin) added to the plasma was between 95 and 106%. Purine derivatives were stable when the processed samples were stored for 7 days at −10°C. The low values of the intra-assay coefficient of variations (2.5–4.6%) and the low values of the detection limits (0.187–0.004 nmol) point to the satisfactory precision and sensitivity of the method. |
| |
Keywords: | Allantoin Uric acid Xanthine Hypoxanthine |
本文献已被 ScienceDirect 等数据库收录! |
|