Post-column enzyme reactors for chemiluminometric detection of glucose, 1,5-anhydroglucitol and 3-hydroxybutyrate in an anion-exchange chromatographic system
Authors:Nobutoshi Kiba  Kazuya Saegusa  Motohisa Furusawa
Institution:Department of Applied Chemistry and Biotechnology, Faculty of Engineering, Yamanashi University, Kofu 400, Japan
Abstract:A liquid chromatographic system consisting of a co-immobilized 3-hydroxybutyrate dehydrogenase-NADH oxidase reactor and an immobilized pyranose oxidase reactor in series and a chemiluminometer was developed for the simultaneous determination of glucose, 1,5-anhydroglucitol and 3-hydroxybutyrate in plasma. The enzymes were immobilized on toresylated poly(vinyl alcohol) beads. Separation was achieved on a TSK gel SAX column (40×4 mm I.D.) with an eluent of 50 mM NaOH containing 30 mM sodium butyrate. The hydrogen peroxide produced was detected by measuring the chemiluminescence emitted on admixing with luminol and potassium hexacyanoferrate(III). The calibration curves were linear from 0.8 to 500 μM (7 ng−4 μg) for glucose, from 0.8 to 400 μM (7 ng−3 μg) for 1,5-anhydroglucitol and from 1 to 700 μM (5 ng−4 μg in a 50-μl injection) for 3-hydroxybutyrate. The sample throughput was four per hour. The reactors were stable for at least ten days.
Keywords:Glucose  1  5-Anhydroglucitol  3-Hydroxybutyrate