Effects of particle characteristics on magnetic immunoassay in a thin channel |
| |
Authors: | Tsai H Y Hsieh Y C Su Y M Chan J R Chang Y C Fuh C Bor |
| |
Institution: | Department of Applied Chemistry, Chung Shan Medical University, Taichung 402, Taiwan. |
| |
Abstract: | The effects of size and porosity of particles on magnetic immunoassay in a thin channel were studied. Experimental parameters were investigated and compared using a model immunoassay complex of carcinoembryonic antigen (CEA)/anti-CEA. The rate constant for the affinity reaction between functional particles increased as the size of magnetic nanoparticles (800-80 nm) decreased. The affinity reaction between functional particles had no significant effect on the sizes of microparticles (1.0-4.4 μm) at commonly used thin channel flow-rates of 0.001-0.025 ml/min. Competitive and sandwich reactions of CEA/anti-CEA were studied for CEA detection. Microparticles of different porosities produced similar linear ranges of detection and limits of detection. The limits of detection for CEA were 0.29 pg/ml and 0.21 pg/ml for competitive and sandwich reactions, respectively. The linear ranges of detection were from 0.49 pg/ml to 4.9 ng/ml for both competitive and sandwich reactions. The detection limits were lower, and the linear ranges were wider than those of literature. There was a 9% difference in CEA detection measurements between competitive and sandwich magnetic immunoassay. The measurements of two magnetic immunoassays differed by less than 13% from the ELISA reference measurements. The running time was less than 30 min. Magnetic immunoassay in a thin channel has great potential for biochemical analysis and immunoassay-related applications. |
| |