The impact of imperfect frame deployment and rotational orientation on stress within the prosthetic leaflets during transcatheter aortic valve implantation
Institution:1. Computational Engineering and Design Group, Faculty of Engineering and the Environment, University of Southampton, Southampton, UK;2. Southampton University Hospitals NHS Trust, Southampton, UK;3. Faculty of Medicine, University of Southampton, Southampton, UK;1. Department of Surgery, University of California San Francisco and San Francisco VA Medical Centers, San Francisco, Calif;2. Division of Cardiovascular Surgery, St Paul''s Hospital and Vancouver General Hospital, Vancouver, British Columbia, Canada;3. Division of Cardiology, University of Washington, Seattle, Wash;1. Dip. di Ingegneria Industriale e dell?Informazione, Università di Pavia, Pavia, Italy;2. ICSA Policlinico Sant?Ambrogio, Milano, Italy;3. Ospedale Cisanello, Pisa, Italy;4. Dip. di Ingegneria Civile e Architettura, Università di Pavia, Pavia, Italy;1. MMDRG, Biomechanics Research Centre, Biomedical Engineering, College of Engineering & Informatics, National University of Ireland Galway, Ireland;2. Wallace H. Coulter Department of Biomedical Engineering, Georgia Institute of Technology, USA;1. Research Institute for Complex Issues of Cardiovascular Diseases, Sosnovy Blvd. 6, Kemerovo 650002, Russian Federation;2. Department of Oncology, CRUK/MRC Oxford Institute for Radiation Oncology, University of Oxford, Roosevelt Drive, Oxford OX3 7DQ, UK;3. Department of Biomedical Technologies, Bauman Moscow State Technical University, Baumanskaya 2-ya Str. 5, Moscow 105005, Russian Federation;4. National Research Tomsk Polytechnic University, Lenina Ave. 30, Tomsk 634050, Russian Federation
Abstract:TAVI devices are manufactured with cylindrical frames. However, the frames are rarely cylindrical post-deployment since deformation due to localised under expansion can be induced by calcified material on the native valve leaflets exerting irregular forces upon the frame. Consequently, the leaflets within a deformed TAVI device may undergo elevated stress during operation, which may lead to premature device failure.Using computational analysis a complete TAVI device model was simulated undergoing deployment into an aortic root model derived from CT data for a patient with severe calcific aortic stenosis, followed by a pressure simulated cardiac cycle. The complete analysis was performed eight times, each with the device at a different rotational orientation relative to the native valve, with an increment spacing of 15°.The TAVI device frames consistently featured significant distortions associated with bulky calcified material at the base of the non-coronary sinus. It was found that the average von Mises stress in the prosthetic valves was only increased in one of the cases relative to an idealised device. However, the maximum von Mises stress in the prosthetic valves was elevated in the majority of the cases.Furthermore, it was found that there were preferable orientations to deploy the prosthetic device, in this case, when the prosthetic leaflets were aligned with the native leaflets. As device orientation deviated from this orientation, the stresses in the valve increased because the distance between the prosthetic commissures decreased. This potentially could represent a sufficient increase in stress to induce variation in device lifespan.
Keywords:TAVI  TAVR  FEA  Leaflet  Simulation
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号