首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Study of probe-sample distance for biomedical spectra measurement
Authors:Bowen Wang  Shuzhen Fan  Lei Li  Cong Wang
Institution:1. Medical Electronics and Information Lab., Department of Electrical & Electronic Engineering, Yonsei University, Seoul, South Korea
Abstract:

Background

There have been many studies that utilize the bio-impedance measurement method to analyze the movements of the upper and lower limbs. A fixed electrical current flows into the limbs through four standard disposable electrodes in this method. The current flows in the muscles and blood vessels, which have relatively low resistivity levels in the human body. This method is used to measure bio-impedance changes following volume changes of muscles and blood vessels around a knee joint. The result of the bio-impedance changes is used to evaluate the movements. However, the method using the standard disposable electrodes has a restriction related to its low bio-impedance changes: the standard disposable electrodes are only able to measure bio-impedance from a limited part of a muscle. Moreover, it is impossible to use continuously, as the electrodes are designed to be disposable. This paper describes a conductive fabric sensor (CFS) using a bio-impedance measurement method and determines the optimum configuration of the sensor for estimating knee joint movements.

Methods

The upper side of subjects' lower limbs was divided into two areas and the lower side of subjects' lower limbs was divided into three areas. The spots were matched and 6 pairs were selected. Subjects were composed of 15 males (age: 30.7 ± 5.3, weight: 69.8 ± 4.2 kg, and height: 173.5 ± 2.8 cm) with no known problems with their knee joints. Bio-impedance changes according to knee joint flexion/extension assessments were calculated and compared with bio-impedance changes by an ankle joint flexion/extension test (SNR I) and a hip joint flexion/extension test (SNR II).

Results

The bio-impedance changes of the knee joint flexion/extension assessment were 35.4 ± 20.0 Ω on the (1, 5) pair. SNR I was 3.8 ± 8.4 and SNR II was 6.6 ± 7.9 on the (1, 5) pair.

Conclusions

The optimum conductive fabric sensor configuration for evaluating knee joint movements were represented by the (1, 5) pair.
Keywords:
本文献已被 SpringerLink 等数据库收录!