In vitro simulation and quantification of wear within the patellofemoral joint replacement
Authors:Ellison P  Barton D C  Esler C  Shaw D L  Stone M H  Fisher J
Institution:

aInstitute of Medical and Biological Engineering, University of Leeds, Leeds LS9 2JT, UK

bDepartment of Orthopaedic Surgery, The Glenfield Hospital, Leicester, UK

cDepartment of Orthopaedic Surgery, Bradford Royal Infirmary, UK

dDepartment of Orthopaedic Surgery, Leeds General Infirmary, UK

Abstract:Quantification of the wear rate in vitro is now considered an essential step in the development of a new joint replacement prior to clinical trials. However, little research exists around in vitro simulation of wear in the patellofemoral joint (PFJ) despite over 200,000 being implanted annually within the European Union. A method to simulate wear in the laboratory using four input degrees of freedom within the PFJ of total knee replacement (TKR) has been developed. Wear simulation was validated through comparison of functional kinematics and patellar surface damage modes produced in vitro to clinical outcomes. The technique has been shown to replicate the prescribed in vivo kinematics in a reproducible and repeatable manner. The wear scar areas were similar to those found in vivo. However, geometrical measurements of wear were not reliable due to creep and geometry changes. As has been found previously with tibial inserts, geometrical determination of wear volume was not found to be an effective method of comparing wear from simulators and retrievals. Change in volume calculated gravimetrically was seen to be the most repeatable measure of patellar wear in vitro.
Keywords:Patella  Replacement  Wear  Knee
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号