Liquid scintillation counting of radioactive monomeric and polymeric carbohydrates on paper chromatograms |
| |
Authors: | P A Sandford P R Watson |
| |
Affiliation: | Department of Ophthalmology, University of Washington School of Medicine, Seattle, Washington 98195 USA |
| |
Abstract: | A simple, improved scintillation counting procedure was developed for the assay of radioactive mono- and polysaccharides on paper chromatograms. Segments of chromatograms are placed in scintillation vials and soaked in water to completely elute the carbohydrate before addition of Aquasol, a xylene-based scintillation fluid. The resulting water-Aquasol solution is counted in a liquid scintillation counter. Evaluation of numerous experimental variables revealed optimal conditions for complete elution of mono- and polysaccharides with water before counting in Aquasol.The water elution-Aquasol procedure allows water-soluble substances (14C- and 3H-labeled) on paper to be assayed with increased accuracy and sensitivity (three- to fivefold improvement in counting efficiency of tritiated samples). The simplicity of the procedure allows entire radiochromatograms to be assayed readily. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|