Mediated, amperometric biosensor for glucose-6-phosphate monitoring based on entrapped glucose-6-phosphate dehydrogenase, Mg2+ ions, tetracyanoquinodimethane, and nicotinamide adenine dinucleotide phosphate in carbon paste
Authors:Bassi A S  Tang D  Bergougnou M A
Institution:Department of Chemical and Biochemical Engineering, Faculty of Engineering Science, University of Western Ontario, London, Ontario, N6A 5B9, Canada.abassi@julian.uwo.ca
Abstract:In this study, an amperometric carbon paste biosensor is developed for glucose-6-phosphate (G6P) monitoring which is based on entrapped Mg2+ ions, G6P dehydrogenase, NADP+ polyethylenimine (PEI) and the electroactive mediator, tetracyanoquinodimethane (TCNQ). The calibration line had a slope of 1.55 x 10(-5) A. M-1 with a correlation coefficient of 0.9965. The limit of detection (defined as three times the standard deviation of the response of the electrode to blank phosphate buffer injections (noise)) of the G6P biosensor was 5.0 x 10(-5) M. The application of this biosensor for monitoring G6P in human blood using the standard addition method is also demonstrated. A two-parameter empirical equation which adequately describes the deactivation of the biosensor steady-state response with time is also proposed.