Quantitative determination of Aconitum alkaloids in blood and urine samples by high-performance liquid chromatography |
| |
Authors: | Wang Zhao Hong Guo Dean He Yi Hu Chunhua Zhang Jizong |
| |
Affiliation: | School of Pharmaceutical Sciences, Peking University, Beijing 100083, People's Republic of China. |
| |
Abstract: | An HPLC method has been developed for the simultaneous determination of the toxic Aconitum alkaloids, aconitine, mesaconitine and hypaconitine in blood and urine samples. The samples were initially subjected to solid phase extraction using Oasis MCX cartridges, and the alkaloids were separated on an XTerra RP18 column, gradient-eluted with acetonitrile: ammonium hydrogen carbonate buffer. Calibration curves were linear in the range 2.75-550 ng for aconitine and hypaconitine, and 3-600 ng for mesaconitine: the limit of detection was 0.1 ng (signal-to-noise ratio of 3) for each alkaloid. The described analysis proved to be sensitive, rapid and economical, and will be applied in the identification and determination of these alkaloids in forensic and therapeutic drug monitoring. |
| |
Keywords: | High‐performance liquid chromatography solid phase extraction quantitative determination Aconitum alkaloids |
本文献已被 PubMed 等数据库收录! |
|