首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Gas chromatographic method for the simultaneous determination of hydralazine and its acetylated metabolite in serum using a nitrogen-selective detector
Authors:Helle Riis Angelo  Jytte Molin Christensen  Mogens Kristensen  Allan McNair
Institution:The Department of Clinical Chemistry, Bispebjerg Hospital, DK-2400 Copenhagen Denmark;Department C, Diakonissestiftelsen, Copenhagen Denmark
Abstract:A relatively simple gas chromatographic method has been developed for the quantitative determination of hydralazine simultaneously with its acetylated metabolite, 3-methyl-s-triazolo3,4-α]phthalazine (MTP). The proteins were removed by means of sulfosalicylic acid and Sure-Sep®. On treatment with formic acid, hydralazine and its internal standard were converted into their formylated derivatives. These derivatives, MTP and its internal standard were extracted with toluene and determined by gas chromatography with a nitrogen-selective detector. The lower limits of detection for hydralazine and MTP were 0.13 and 0.27 μmol/l, respectively.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏