Investigations of the performance of a high-performance liquid chromatography system with an electrochemical detector |
| |
Authors: | Gary Wenk Robert Greenland |
| |
Affiliation: | University of Cincinnati, Institute of Environmental Health, Kettering Laboratory, Cincinnati, OH 45267 U.S.A. |
| |
Abstract: | A procedure for the determination of norepinephrine and dopamine, based on high-performance liquid chromatography, is evaluated using an electrochemical detector system. The use of an inorganic mobile phase to provide resolution of low retention amines and extend column life is discussed. A high degree of correlation between estimations of endogenous catecholamine levels is reported using both electrochemical and fluorometric detector systems. Inter-assay reproducibility of the extraction method, and sensitivity and linearity of response of the electrochemical detector system are shown to be consistent across trials. The system described is determined to be accurate, sensitive, and reliable over time. |
| |