首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Sensitive HPLC-APCI-MS method for the determination of cyclovirobuxine D in human plasma
Authors:Ding Li  Hu Jingjing  Jiang Meng  Xiong Ningning
Institution:Department of Pharmaceutical Analysis, China Pharmaceutical University, 24 Tongjiaxiang, Nanjing 210009, China.
Abstract:A sensitive high performance liquid chromatography-atmospheric pressure chemical ionisation-mass spectrometry (HPLC-APCI-MS) assay for determination of cyclovirobuxine D (CVB-D) in human plasma using mirtazapine as internal standard (I.S.) was established. After adjustment to a basic pH with sodium hydroxide, plasma was extracted by ethyl acetate and separated by high performance liquid chromatography (HPLC) on a reversed-phase C(18) column with a mobile phase of 30 mM ammonium acetate buffer solution containing 1% formic acid-methanol (48:52, v/v). CVB-D was determined with atmospheric pressure chemical ionisation-mass spectrometry (APCI-MS). HPLC-APCI-MS was performed in the selected-ion monitoring (SIM) mode using target ions at M+H](+)m/z 403.4 for CVB-D and M+H](+)m/z 266.2 for I.S. Calibration curves were linear over the range 10.11-4044 pg/ml. The lower limit of quantification was 10.11 pg/ml. The intra- and inter-run variability values were less than 9.5 and 12.4%, respectively. The mean plasma extraction recovery of CVB-D was in the range of 85.3-92.8%. The method was successfully applied to determine the plasma concentrations of CVB-D in Chinese volunteers.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏