Reductive determination of hydrogen peroxide with MWCNTs-Pd nanoparticles on a modified glassy carbon electrode |
| |
Authors: | You Jung-Min Jeong Yu Na Ahmed Mohammad Shamsuddin Kim Seul Ki Choi Hyun Chul Jeon Seungwon |
| |
Affiliation: | Department of Chemistry and Institute of Basic Science, Chonnam National University, Gwangju 500-757, Republic of Korea. |
| |
Abstract: | This paper introduces the use of multi walled carbon nanotubes (MWCNTs) with palladium (Pd) nanoparticles in the electrocatalytic reduction of hydrogen peroxide (H(2)O(2)). We have developed and characterized a biosensor for H(2)O(2) based on Nafion(?) coated MWCNTs-Pd nanoparticles on a glassy carbon electrode (GCE). The Nafion(?)/MWCNTs-Pd/GCE electrode was easily prepared in a rapid and simple procedure, and its application improves sensitive determination of H(2)O(2). Characterization of the MWCNTs-Pd nanoparticle film was performed with transmission electron microscopy (TEM), Raman, and X-ray photoelectron spectroscopy (XPS). Cyclic voltammetry (CV) and amperometry (at an applied potential of -0.2V) measurements were used to study and optimize performance of the resulting peroxide biosensor. The proposed H(2)O(2) biosensor exhibited a wide linear range from 1.0 μM to 10 mM and a low detection limit of 0.3 μM (S/N=3), with a fast response time within 10s. Therefore, this biosensor could be a good candidate for H(2)O(2) analysis. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|