Optimisation of cardiac resynchronization therapy in clinical practice during exercise
Authors:M M D Molenaar  B Oude Velthuis  M F Scholten  J Y Stevenhagen  W A Wesselink  J M van Opstal
Institution:1. Thoraxcentrum Twente, Medisch Spectrum Twente, Haaksbergerstraat 55, 7513 ER, Enschede, the Netherlands
2. Technical Medicine, Twente University, Enschede, the Netherlands
3. Clinical Team, Edwards Lifesciences BMEYE, Amsterdam, the Netherlands
Abstract:

Aims

Although cardiac resynchronisation therapy (CRT) is an established treatment to improve cardiac function, a significant amount of patients do not experience noticeable improvement in their cardiac function. Optimal timing of the delay between atrial and ventricular pacing pulses (AV delay) is of major importance for effective CRT treatment and this optimum may differ between resting and exercise conditions. In this study the feasibility of haemodynamic measurements by the non-invasive finger plethysmographic method (Nexfin) was used to optimise the AV delay during exercise.

Methods and results

Thirty-one patients implanted with a CRT device in the last 4 years participated in the study. During rest and in exercise, stroke volume (SV) was measured using the Nexfin device for several AV delays. The optimal AV delay at rest and in exercise was determined using the least squares estimates (LSE) method. Optimisation created a clinically significant improvement in SV of 10 %. The relation between HR and the optimal AV delay was patient dependent.

Conclusion

A potential increase in SV of 10 % can be achieved using Nexfin for optimisation of AV delay during exercise. A considerable number of patients showed benefit with lengthening of the AV delay during exercise.
Keywords:Cardiac resynchronization therapy  Atrio-ventricular delay  Optimisation
本文献已被 SpringerLink 等数据库收录!