首页 | 本学科首页   官方微博 | 高级检索  
     


Automated analysis of uronic acids by high-performance liquid chromatography with photometric and fluorimetric postcolumn labelling using 2-cyanoacetamide
Authors:S Honda  S Suzuki  M Takahashi  K Kakehi  S Ganno
Affiliation:1. Faculty of Pharmaceutical Sciences, Kinki University, 3-4-1 Kowakae, Higashi-osaka, Japan;1. Naka Works, Hitachi Ltd., Katsuta, Ibaraki, Japan
Abstract:A convenient method for the rapid and sensitive automated analysis of uronic acids, based on high-performance anion-exchange chromatography on a Hitachi 2633 column and photometric as well as fluorimetric postcolumn labeling with 2-cyanoacetamide, has been developed. This method allows the simultaneous determination of 1-1000 nmol of D-mannuronic and D-galacturonic acids and 5-1000 nmol of L-iduronic and D-glucuronic acids in approx 70 min with high precision by photometric monitoring. In fluorimetric monitoring the linearity range was 1-1000 nmol for all these uronic acids, but reproducibility was rather low at the lowest limit of linearity. Application of this method to the analysis of component uronic acids in some polyuronides has suggested the inadequacy of generally accepted conditions for hydrolysis.
Keywords:automated analysis  uronic acids  high-performance anion-exchange chromatography  borate complexes  photometric and fluorimetric postcolumn labeling  2-cyanoacetamide
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号