首页 | 本学科首页   官方微博 | 高级检索  
   检索      


GraFix: sample preparation for single-particle electron cryomicroscopy
Authors:Kastner Berthold  Fischer Niels  Golas Monika Mariola  Sander Bjoern  Dube Prakash  Boehringer Daniel  Hartmuth Klaus  Deckert Jochen  Hauer Florian  Wolf Elmar  Uchtenhagen Hannes  Urlaub Henning  Herzog Franz  Peters Jan Michael  Poerschke Dietmar  Lührmann Reinhard  Stark Holger
Institution:Department of Cellular Biochemsitry, Max Planck Institute for Biophysical Chemistry, Am Fassberg 11, 37077 G?ttingen, Germany. bkastne@gwdg.de
Abstract:We developed a method, named GraFix, that considerably improves sample quality for structure determination by single-particle electron cryomicroscopy (cryo-EM). GraFix uses a glycerol gradient centrifugation step in which the complexes are centrifuged into an increasing concentration of a chemical fixation reagent to prevent aggregation and to stabilize individual macromolecules. The method can be used to prepare samples for negative-stain, cryo-negative-stain and, particularly, unstained cryo-EM.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏