The accuracy of joint surface models constructed from data obtained with an electromagnetic tracking device |
| |
Authors: | van Ruijven L J Beek M Donker E van Eijden T M |
| |
Affiliation: | Department of Functional Anatomy, Academic Centre for Dentistry Amsterdam (ACTA), Netherlands. l.j.vanruijven@amc.uva.nl |
| |
Abstract: | Electromagnetic tracking devices are widely used in biomechanics. In this article a method is evaluated to construct models of articular surfaces using an electromagnetic tracking device. First, the accuracy of the space tracker was examined and optimised. Then, from several joint surfaces random points were measured and eighth degree polynomials were fitted to these measurements. To check if the fit converged well, plots of cross sections of the model with corresponding data points were examined. The accuracy of the models was determined by comparing them with computed tomography data and by reproducibility tests. All the fits converged well to the data. The root mean square (RMS) error of the models varied from 0.07 to 0. 18mm, and was proportional to the size and complexity of the surface. This was mainly due to systematic errors made by the space tracker, which were also proportional to the size and complexity of the surface. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|