首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Intravascular ultrasound assessment of culotte stent deployment for the treatment of stenoses at major coronary bifurcations
Authors:Robinson Nicholas M  Balcon Raphael  Layton Clive A  Mills Peter G  Timmis Adam D  Rothman Martin T
Institution:The London Chest Hospital, London, UK.
Abstract:BACKGROUND: The mechanism for the disappointing late outcome following stenting of bifurcation lesions is unclear. This prospective observational study aims to evaluate culotte stent deployment and dimensions with intravascular ultrasound (IVUS). PATIENTS AND METHODS: Patients with bifurcation stenoses were treated using two stents in a culotte configuration. After optimizing the angiographic appearance of both stents, IVUS was used to evaluate both limbs of the culotte. The main outcome measures were cross-sectional area (CSA) and minimal lumen diameter (MLD) assessed by IVUS. RESULTS: Within the culotte stent, the final mean CSA in the main limb was 6.1 mm(2) (97% of reference) and in the side-limb was 5.9 mm(2) (97% of reference). However, in each case, the minimum CSA and IVUS MLD of both limbs was at the bifurcation point. For all patients, the final mean CSA at the bifurcation point of the main limb was 4.3 mm(2) (70% of main stent) and of the side-limb was 4.4 mm(2) (75% of side stent). The IVUS MLD at the bifurcation point of the main limb was 2.1 mm (78% of main stent) and of the side-limb was 2.1 mm (84% of the side stent). Importantly, this significant residual stenosis was not detectable with quantitative coronary angiography. CONCLUSIONS: IVUS evaluation of culotte stents is feasible. The minimum IVUS CSA and MLD of both limbs of the culotte stent is at the bifurcation point. Despite an optimal angiographic appearance a significant residual stenosis was noted with IVUS at each bifurcation point.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号