Determination of glutathione and glutathione disulfide in hepatocytes by liquid chromatography with an electrode modified with functionalized carbon nanotubes |
| |
Authors: | Zhang Wen Wan Fangli Zhu Wei Xu Haihong Ye Xiyun Cheng Rongyi Jin Li-Tong |
| |
Affiliation: | Department of Chemistry, East China Normal University, Shanghai 200062, PR China. wenzhang26@163.com |
| |
Abstract: | Glutathione (GSH) and glutathione disulfide (GSSG) are important thiols, which provide defence against oxidative stress by scavenging free radicals or causing the reduction of hydrogen peroxide. The ratio GSH/GSSG is often used as a sensitive index of oxidative stress in vivo. In this paper, a direct electrochemical method using an electrode modified with functionalized carbon nanotubes as electrochemical detector (ED) for liquid chromatography (LC) was described. The electrochemical behaviors of GSH and GSSG on this modified electrode were investigated by cyclic voltammetry and it was found that the functionalized carbon nanotubes exhibited efficiently electrocatalysis on the current responses of GSH and GSSG. In LC-ED, both of the analytes showed good and stable current responses. The detection limit of GSH was 0.2 pmol on column and that of GSSG was 1.2 pmol on column, which were low enough for the analysis of real small samples. The method was sensitive enough to detect difference in concentration of GSH and GSSG in hepatocytes from animals with and without introduction of oxidation stress by glucose or hydrogenperoxide. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|