首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Low-level quantification of melamine and cyanuric acid in limited samples of rat serum by UPLC-electrospray tandem mass spectrometry
Authors:Jacob Cristina C  Gamboa da Costa Gonçalo
Institution:Division of Biochemical Toxicology, US FDA National Center for Toxicological Research, Jefferson, AR 72079, USA.
Abstract:This paper reports the development and validation of a methodology for the low-level quantification of melamine and cyanuric acid in limited samples of rat serum. The methodology, based upon ion-exchange solid phase extraction (SPE) and ultra-performance liquid chromatography (UPLC) coupled with electrospray tandem mass spectrometry (MS/MS) in multiple reaction monitoring (MRM) mode, relies on the use of stable isotope-labeled internal standards and requires only 15 μL samples of serum. The method provides a recovery of 80-110% of melamine with a signal suppression of ca. 55%, and a recovery of 50-90% of cyanuric acid with a signal suppression ca. 40-60%, affording lower limits of quantification (LLOQ) for melamine or cyanuric acid of, respectively, 5 ppb (mean accuracy 109%; CV=4.9%) and 10 ppb (mean accuracy 96%; CV=8.6%). The small sample requirements, excellent sensitivity, accuracy and precision, and high-throughput (5 min of instrument run time) make this methodology optimal for toxicokinetic or exposure assessments studies.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号