首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of selenium in serum by FI-HG-AAS and calculation of dietary intake
Authors:Fangshi Li  E. Rossipal  Dusanka Micetic-Turk
Affiliation:(1) Department of Applied Chemistry, Nanjing University of Chemical Technology, 210009 Nanjing, People’s Republic of China;(2) Department of Pediatrics., Karl-Franzens University Graz, A-8010 Graz, Austria;(3) Department of Pediatrics, Teaching Hospital, Maribor, Slovenia
Abstract:A method was developed for the determination of selenium concentration in serum by flow injection-hydride generation-atomic absorption spectrometry (FI-HG-AAS) following microwave digestion of serum samples and reduction of selenate to selenite. The detection limit of the method was 0.3 microg Se/L and the characteristic concentration, corresponding to the 0.0044 absorbance signal, was 0.12 microg Se/L. The results from the analysis of two Seronorm standard reference materials showed good agreement with the certified values. The method was then used to analyze selenium in sera of Austrian and Slovenian people for the calculation of dietary intakes. The selenium concentrations in sera of mothers at delivery, their neonates, and the male and female adults were 71+/-14, 42+/-6, 75+/-21, and 65+/-16 microg/L for the Austrians and 62+/-15, 34+/-7, 70+/-12, and 66+/-15 microg/L for the Slovenians. The dietary intakes of selenium of the mothers and the male and the female adults were calculated as 52, 37, and 46 microg/d for the Austrians and 45, 38, and 32 microg/d for the Slovenians.
Keywords:Selenium  Austria  Slovenia  serum  dietary intake  FI-HG-AAS  mircrowave digestion
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号