Low-volume,high-sensitivity assay for cadmium in blood and urine using conventional atomic absorption spectrophotometry
Authors:Cerny Elizabeth A  Bhattacharyya Maryka H
Institution:Biosciences Division, Argonne National Laboratory, Argonne, IL 60439, USA. cerny@anl.gov
Abstract:An assay for cadmium in whole blood and urine using deuterium background-correction electrothermal atomic absorption spectroscopy (D(2)-ETAAS) was developed. Cadmium (in a 1- to 2-ml sample) was bound to 15 mg anion-exchange resin, interfering ions were removed in a 2-ml Bio-Spin column, and cadmium was extracted into 100 microl 1M nitric acid for analysis. Cadmium in the sample extract was concentrated 7-fold for blood and 10-fold for urine over the starting material. These steps produced cadmium atomic absorption traces with high signal to background ratios and allowed analysis against aqueous standards. At approximately 0.1 ng Cd/ml, mean intra- and interassay coefficients of variation were 11-12%. Cadmium recovery for 0.1 to 0.6 ng added cadmium was 107+/-4% for blood and 94+/-4% for urine (mean+/-SE, n=3). The mean detection limit (mean + 3 x SD of blank) was 0.008 ng/ml for blood and 0.003 ng/ml for urine. Samples from "unexposed" animals including humans ranged from 0.051+/-0.000 to 0.229+/-0.035 ng/ml. Values were approximately 10-fold lower than those obtained by the method of Stoeppler and Brandt using Zeeman background-correction ETAAS. This new high-sensitivity, low-volume assay will be useful for epidemiological studies, even those involving children, and will provide a means to help determine the contribution of cadmium to disease incidence in the general population.
Keywords:Cadmium  Assay  Blood  Urine  Atomic absorption spectrophotometry
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号