Comparison of Dixon Sequences for Estimation of Percent Breast Fibroglandular Tissue |
| |
Authors: | Araminta E W Ledger Erica D Scurr Julie Hughes Alison Macdonald Toni Wallace Karen Thomas Robin Wilson Martin O Leach Maria A Schmidt |
| |
Institution: | 1. CR-UK Cancer Imaging Centre, The Institute of Cancer Research and Royal Marsden NHS Foundation Trust, London, United Kingdom;2. Department of Radiology, Royal Marsden NHS Foundation Trust, London, United Kingdom;3. Clinical Research and Development, Royal Marsden NHS Foundation Trust, London, United Kingdom;University of Chicago, UNITED STATES |
| |
Abstract: | ObjectivesTo evaluate sources of error in the Magnetic Resonance Imaging (MRI) measurement of percent fibroglandular tissue (%FGT) using two-point Dixon sequences for fat-water separation.MethodsTen female volunteers (median age: 31 yrs, range: 23–50 yrs) gave informed consent following Research Ethics Committee approval. Each volunteer was scanned twice following repositioning to enable an estimation of measurement repeatability from high-resolution gradient-echo (GRE) proton-density (PD)-weighted Dixon sequences. Differences in measures of %FGT attributable to resolution, T1 weighting and sequence type were assessed by comparison of this Dixon sequence with low-resolution GRE PD-weighted Dixon data, and against gradient-echo (GRE) or spin-echo (SE) based T1-weighted Dixon datasets, respectively.Results%FGT measurement from high-resolution PD-weighted Dixon sequences had a coefficient of repeatability of ±4.3%. There was no significant difference in %FGT between high-resolution and low-resolution PD-weighted data. Values of %FGT from GRE and SE T1-weighted data were strongly correlated with that derived from PD-weighted data (r = 0.995 and 0.96, respectively). However, both sequences exhibited higher mean %FGT by 2.9% (p < 0.0001) and 12.6% (p < 0.0001), respectively, in comparison with PD-weighted data; the increase in %FGT from the SE T1-weighted sequence was significantly larger at lower breast densities.ConclusionAlthough measurement of %FGT at low resolution is feasible, T1 weighting and sequence type impact on the accuracy of Dixon-based %FGT measurements; Dixon MRI protocols for %FGT measurement should be carefully considered, particularly for longitudinal or multi-centre studies. |
| |
Keywords: | |
|
|