Rapid high-performance liquid chromatographic determination of serum gabapentin
Authors:Qibo Jiang  Shuguang Li
Abstract:Gabapentin (GBP) is a new antiepileptic drug approved for clinical treatment of partial seizures in the USA. Serum GBP concentrations in 283 patients were studied using high-performance liquid chromatography with fluorescence detection. The standard curves were linear over a range of 60 ng to 15 μg/ml. The coefficient of variations were 3.4 to 8.8% and 1.4 to 9.8% for intra- and inter-assay studies, respectively. The lower limit of quantitation was 10 ng/ml. Of the 283 patients studied, 72.5% had GBP levels between 2 and 10 μg/ml, 14.8% were below 2 μg/ml and 12.7% above 10 μg/ml. The mean±S.E. of GBP in 283 patients was 5.38±0.23 μg/ml. Peak concentrations of more than 15 μg/ml and trough levels as low as 0.1 μg/ml were not uncommon. The method described was rapid, simple, highly sensitive and reproducible. Other antiepileptic drugs and endogenous compounds did not interfere with the assay.