Successful termination of ventricular tachycardia with intrinsic anti-tachycardia pacing
Affiliation:1. Department of Cardiology, Ogaki Municipal Hospital, Ogaki, Japan;2. Department of Clinical Engineering, Ogaki Municipal Hospital, Ogaki, Japan
Abstract:Intrinsic anti-tachycardia pacing (iATP) is a novel automated ATP algorithm that employs post-pacing interval (PPI) to design the next ATP sequence based on an analysis of the prior failed ATP sequence. A patient with hypertrophic cardiomyopathy received an implantable cardioverter-defibrillator (ICD) (Cobalt™ XT DR, Medtronic, Minneapolis, MN, USA) following an episode of syncope due to macro-reentrant ventricular tachycardia (VT) (right bundle branch block configuration, cycle length [CL] 280 ms). The VF zone was set to VTCL <300 ms and iATP therapy was prescribed before and during capacitor charging. The iATP was initiated when VT recurred 3 months later. The first attempt with an assumption of 150 ms propagation time from the pacing site to the VT circuit (9 pulses) could not reset the VT, leaving a PPI of 650 ms. A subsequent attempt involving 20 pulses with an assumption of 250 ms propagation time terminated the VT. Failure to reach the circuit is a major cause of unsuccessful ATP. In this regard, iATP is expected to have theoretical advantages over empirical and traditional ATP therapies. To the best of our knowledge, this is the first intracardiac electrogram illustrating how automated precision ATP terminates VT in a clinical setting.
Keywords:Implantable cardioverter-defibrillator  Intrinsic anti-tachycardia pacing  Ventricular tachycardia