首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Method for the design and analysis of a non-linear anisotropic lower limb prosthetic socket
Institution:1. Michigan State University Mechanical Engineering, 428 S. Shaw Lane, East Lansing, MI 48824, United States;2. Department of Mechanical Engineering, Michigan State University, 428 S. Shaw Lane, East Lansing, MI 48824, United States
Abstract:A simplified computer-based structural analysis procedure has been developed for evaluating the stresses in a lower limb prosthetic socket when subjected to external loads. This technique has been tested by studying a simple linear axisymmetric shell subjected to uniform loads. It was found that the stress and strain distributions obtained are close to those determined by a finite element technique. The method that has been proposed involves an incremental procedure, which can be used to analyse the behaviour of a prosthetic socket of irregular shape made from non-linear anisotropic materials. By employing the proposed procedure to study an example, in which a simplified lower limb prosthetic socket is subject to internal pressure and frictional forces, the effects of the material non-linearity have been found to be significant.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号