首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Miniaturized glucose biosensor modified with a nitric oxide-releasing xerogel microarray
Authors:Oh Bong Kyun  Robbins Mary E  Nablo Brian J  Schoenfisch Mark H
Institution:Department of Chemistry, University of North Carolina at Chapel Hill, Chapel Hill, NC 27599-3290, USA.
Abstract:An enzyme-based glucose biosensor modified to release nitric oxide (NO) via a xerogel microarray is reported. The biosensor design is as follows: (1) glucose oxidase (GOx) is immobilized in a methyltrimethoxysilane (MTMOS) xerogel layer; (2) a blended polyurethane/hydrophilic polyurethane coating prevents enzyme leaching and imparts selectivity for glucose; and (3) micropatterned xerogel lines (5 microm wide) separated by distances of 5 or 20 microm provide NO-release capability. This configuration allows for increased glucose sensitivity relative to sensors modified with NO-releasing xerogel films since significant portions of the sensor surface remain unmodified. Glucose diffusion to the GOx layer is thus less inhibited. The micropatterned NO-releasing biosensors generate sufficient NO levels to reduce both Pseudomonas aeruginosa and platelet adhesion without significantly compromising the enzymatic activity of GOx. The glucose response, linearity and stability of the NO-releasing micropatterned sensors are reported.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏