首页 | 本学科首页   官方微博 | 高级检索  
     


In vivo dosimetry with MOSFETs and GAFCHROMIC films during electron IORT for Accelerated Partial Breast Irradiation
Affiliation:1. Medical Physics Service of the Department of Radiation Oncology, Humanitas Research Hospital, Rozzano, Milan, Italy;2. School of Physics, The University of Sydney, Sydney, Australia;3. Department of Radiation Oncology, Chris O''Brien Lifehouse, Camperdown, Australia;4. Department of Radiation Oncology, Humanitas Research Hospital, Rozzano, Milan, Italy
Abstract:PurposeThe purpose of this study was to compare the delivered dose to the expected intraoperative radiation therapy (IORT) dose with in vivo dosimetry. For IORT using electrons in accelerated partial breast irradiation, this is especially relevant since a high dose is delivered in a single fraction.MethodsFor 47 of breast cancer patients, in vivo dosimetry was performed with MOSFETs and/or GAFCHROMIC EBT2 films. A total dose of 23.33 Gy at dmax was given directly after completing the lumpectomy procedure with electron beams generated with an IORT dedicated mobile accelerator. A protection disk was used to shield the thoracic wall.ResultsThe results of in vivo MOSFET dosimetry for 27 patients and GAFROMIC film dosimetry for 20 patients were analysed. The entry dose for the breast tissue, measured with MOSFETs, (mean value 22.3 Gy, SD 3.4%) agreed within 1.7% with the expected dose (mean value 21.9 Gy). The dose in breast tissue, measured with GAFCHROMIC films (mean value 23.50 Gy) was on average within 0.7% (SD = 3.7%, range −5.5% to 5.6%) of the prescribed dose of 23.33 Gy.ConclusionsThe dose measured with MOSFETs and GAFROMIC EBT2 films agreed well with the expected dose. For both methods, the dose to the thoracic wall, lungs and heart for left sided patents was lower than 2.5 Gy even when 12 MeV was applied. The positioning time of GAFCHROMIC films is negligible and based on our results we recommend its use as a standard tool for patient quality assurance during breast cancer IORT.
Keywords:Electron IORT  Accelerated Partial Breast Irradiation  Quality Assurance
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号