Glucose biosensor based on nano-SiO2 and "unprotected" Pt nanoclusters
Authors:Yang Haipeng  Zhu Yongfa
Affiliation:

aDepartment of Chemistry, Tsinghua University, Beijing 100084, PR China

bCollege of Materials Science and Engineering, Shenzhen University, Shenzhen Key Laboratory of Special Functional Materials, Shenzhen 518060, PR China

Abstract:The “unprotected” Pt nanoclusters (average size 2 nm) mixed with the nanoscale SiO2 particles (average size 13 nm) were used as a glucose oxidase immobilization carrier to fabricate the amperometric glucose biosensor. The bioactivity of glucose oxidase (GOx) immobilized on the composite was maintained and the as-prepared biosensor demonstrated high sensitivity (3.85 μA mM−1) and good stability in glucose solution. The Pt–SiO2 biosensor showed a detection limit of 1.5 μM with a linear range from 0.27 to 4.08 mM. In addition, the biosensor can be operated under wide pH range (pH 4.9–7.5) without great changes in its sensitivity. Cyclic voltammetry measurements showed a mixed controlled electrode reaction.
Keywords:“Unprotected” Pt nanoclusters   Glucose oxidase   Glucose biosensor   Nanoscale SiO2