Quantitation of promethazine and metabolites in urine samples using on-line solid-phase extraction and column-switching |
| |
Authors: | Q Song L Putcha |
| |
Abstract: | A chromatographic method for the quantitation of promethazine (PMZ) and its three metabolites in urine employing on-line solid-phase extraction and column-switching has been developed. The column-switching system described here uses an extraction column for the purification of PMZ and its metabolites from a urine matrix. The extraneous matrix interference was removed by flushing the extraction column with a gradient elution. The analytes of interest were then eluted onto an analytical column for further chromatographic separation using a mobile phase of greater solvent strength. This method is specific and sensitive with a range of 3.75–1400 ng/ml for PMZ and 2.5–1400 ng/ml for the metabolites promethazine sulfoxide, monodesmethyl promethazine sulfoxide and monodesmethyl promethazine. The lower limits of quantitation (LLOQ) were 3.75 ng/ml with less than 6.2% C.V. for PMZ and 2.50 ng/ml with less than 11.5% C.V. for metabolites based on a signal-to-noise ratio of 10:1 or greater. The accuracy and precision were within ±11.8% in bias and not greater than 5.5% C.V. in intra- and inter-assay precision for PMZ and metabolites. Method robustness was investigated using a Plackett–Burman experimental design. The applicability of the analytical method for pharmacokinetic studies in humans is illustrated. |
| |
Keywords: | Promethazine Promethazine sulfoxide Monodesmethylpromethazine sulfoxide Monodesmethyl promethazine |