首页 | 本学科首页   官方微博 | 高级检索  
   检索      


A solid-state,portable instrument for measurement of chlorophyll luminescence induction in plants
Authors:Ulrich Schreiber  Ulrich Schliwa
Institution:(1) Lehrstuhl Botanik I, Universität Würzburg, Mittlerer Dallenbergweg 64, D-87 Würzburg, F.R.G.
Abstract:A newly developed compact instrument is described for the measurement of chlorophyll luminescence induction in plants. The instrument operates with a pulsed light emitting diode (LED) as light source and a photodiode as luminescence detector. A special emitter-detector geometry provides for high irradiance of the sample and efficient collection of luminescence by the detector. With insertion of appropriate filters the same probe is also suited for measuring prompt chlorophyll fluorescence. The instrument shows considerable flexibility with respect to pulse frequency, relative lengths of light/dark intervals and luminescence sampling periods. Due to a selective amplifier system only that part of luminescence is processed which is induced by the individual excitation pulses. By this approach, the problem of ldquoslow phase accumulationrdquo, encountered with conventional phosphoroscopes, is eliminated. Some examples are given for system operation, demonstrating satisfactory performance in measurements with intact leaves and isolated chloroplasts.
Keywords:chlorophyll luminescence  luminometer  phosphoroscope  photosynthesis  delayed fluorescence