Abstract:A fully automated immunoassay of 17beta-estradiol (E2) was performed using anti-E2 monoclonal antibody immobilized on bacterial magnetic particles (AntiE2-BMPs) and alkaline phosphatase-conjugated E2 (ALP-E2). E2 concentration in environmental water samples was evaluated by decrease in luminescence based on competitive reaction. A linear correlation between the luminescence intensity and E2 concentration was obtained between 0.5 and 5 ppb. The minimum detectable concentration of E2 was 20 ppt. All measurement steps were done within 0.5 h. The analysis of environmental water samples by a commercially available ELISA kit and the BMP-based immunoassay gave good correlation plots with a correlation efficient of 0.992. These results suggest that the fully automated system using the BMP-based immunoassay has some advantages in the high rapidity and sensitivity of the measurement. This system will enable us to determine low E2 concentrations without sample condensation.