Biosensor system for continuous flow determination of enzyme activities. II. Simultaneous determination of plural enzyme activities.
Authors:H Okuma  H Takahashi  S Sekimukai  E Watanabe
Affiliation:Research and Development Laboratory, New Japan Radio Co. Ltd., Saitama, Japan.
Abstract:A biosensor system for continuous flow determination of plural enzyme activities was prepared from the combination of two pyruvate sensors, a prereactor and a flow cell. This system was applied to the simultaneous determination of lactic dehydrogenase (LDH) and glutamic-pyruvic transaminase (GPT) activities in the same sample. These enzyme activities can be determined by measuring pyruvate produced by the enzyme reactions as follows. The amount of pyruvic acid can also be determined from the amount of oxygen consumed upon oxidation of pyruvic acid by pyruvate oxidase. (Formula: see text). Therefore, both of the detectors for the determination of lactic dehydrogenase and glutamic-pyruvic transaminase activities were prepared from the combination of a pyruvate oxidase membrane and an oxygen electrode. Pyruvate oxidase was covalently immobilized on a membrane prepared from cellulose triacetate. A linear relation was obtained between the output current and LDH or GPT activities in the range of 50 to 3,600 IU l-1 or 6 to 1,000 IU l-1, respectively. Each assay of these enzyme activities was completed within 15 min. The results obtained had a precision of ca. 4%. The sensor was stable for more than 25 days at 5 degrees C.