首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Comparison of the conventional culture,the manual fluorescent MGIT system and the automated fluorescent MGIT 960 culture system for the detection of <Emphasis Type="Italic">Mycobacterium avium</Emphasis> ssp. <Emphasis Type="Italic">avium</Emphasis> in tissues of naturally infected hens
Authors:E J Shitaye  V Beran  J Svobodová  M Morávková  V Babák  I Pavlík
Institution:(1) Veterinary Research Institute, 621 00 Brno, Czech Republic;(2) Regional Institute of Public Health, 613 00 Brno, Czech Republic
Abstract:Different methods for the detection of Mycobacterium avium ssp. avium (MAA) in naturally infected hens were compared. They included the conventional culture method (solid Herrold’s and Stonebrink media and liquid Sula medium) and newly developed liquid culture systems, the manual mycobacteria growth indicator tube (M-MGIT) and the fully automated BACTEC MGIT 960 system (A-MGIT). 152 tissues originating from 15 naturally infected hens have been processed. The overall detection rates (percentage of positive cultures from the number of positive cultures determined by all the methods together) were 60, 70 and 76 % for the conventional media, M-MGIT and A-MGIT systems, respectively, the mean time of mycobacteria detection being 32.6, 17.6 and 14.6 d, respectively. The lowest contamination rate (2.0 %) was found in A-MGIT compared with M-MGIT (4.6 %) and conventional media (10.4 %).
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏