首页 | 本学科首页   官方微博 | 高级检索  
     


Modification of an automated liquid-handling system for reagent-jet, nanoliter-level dispensing
Authors:Hicks J S  Harker B W  Beattie K L  Doktycz M J
Affiliation:Oak Ridge National Laboratory, Oak Ridge, TN, USA.
Abstract:Reducing the scale of biochemical reactions is becoming commonplace. Examples include the screening of large libraries of chemical compounds or gene sequences. These applications demand the ability to transfer sub-microliter volumes of fluid. To this end, we have modified a Hamilton MICROLAB 2200 with high-speed solenoids and a liquid pressurization system to modulate volume delivery down to the nanoliter level. Additional modifications include the use of sapphire-tipped dispensing nozzles and a high-resolution substage to assist in the construction of DNA microarrays. Techniques for characterizing the dispensed volume are presented.
Keywords:
本文献已被 PubMed 等数据库收录!