Spectrophotometric determination and removal of unchelated europium ions from solutions containing Eu-diethylenetriaminepentaacetic acid chelate–peptide conjugates |
| |
Authors: | NGR Dayan Elshan Renata Patek Josef Vagner Eugene A Mash |
| |
Institution: | 1. Department of Chemistry and Biochemistry, University of Arizona, Tucson, AZ 85721, USA;2. The Bio5 Institute, University of Arizona, Tucson, AZ 85721, USA |
| |
Abstract: | Europium chelates conjugated with peptide ligands are routinely used as probes for conducting in vitro binding experiments. The presence of unchelated Eu ions in these formulations gives high background luminescence and can lead to poor results in binding assays. In our experience, the reported methods for purification of these probes do not achieve adequate removal of unchelated metal ions in a reliable manner. In this work, a xylenol orange-based assay for the quantification of unchelated metal ions was streamlined and used to determine levels of metal ion contamination as well as the success of metal ion removal on attempted purification. We compared the use of Empore chelating disks and Chelex 100 resin for the selective removal of unchelated Eu ions from several Eu-diethylenetriaminepentaacetic acid chelate–peptide conjugates. Both purification methods gave complete and selective removal of the contaminant metal ions. However, Empore chelating disks were found to give much higher recoveries of the probes under the conditions used. Related to the issue of probe recovery, we also describe a significantly more efficient method for the synthesis of one such probe using Rink amide AM resin in place of Tentagel S resin. |
| |
Keywords: | Europium Chelate Xylenol orange Empore Chelex |
本文献已被 ScienceDirect 等数据库收录! |
|