Discrepancy between radioimmunoassay and high performance liquid chromatography tandem-mass spectrometry for the analysis of androstenedione
Authors:Stephen J Bruce  François ReyAlexandre Béguin  Christine BerthodDominique Werner  Hugues Henry
Institution:Clinical Chemistry, University Hospital of Lausanne, CHUV (Centre Hospitalier Universitaire Vaudois), Route du Bugnon 46, 1011 Lausanne, Switzerland
Abstract:The discrepancy of results for the quantification of androstenedione in human serum between a radioimmunoassay (RIA) method and high performance liquid chromatography tandem-mass spectrometry (LC–MS/MS) was investigated. RIA overestimated concentrations compared to LC–MS/MS on 59 clinical samples (RIA = 1.79 × LC–MS/MS + 0.94). RIA kit and LC–MS/MS calibrants were also determined by both methods. The RIA performed with improved accuracy on the calibrants (RIA = 1.35 × LC–MS/MS − 0.28). Lipid, protein, electrolyte content, and pH of the two sets of calibrants were further investigated. The RIA calibrants contained little lipid material, while the LC–MS/MS calibrant material contained the same levels expected in normal serum/plasma. The pH and sex hormone binding globulin (SHBG) values were different between the RIA calibrants and the LC–MS/MS calibrant material (SHBG, 31 ± 2 and 38 ± 2 nmol/l; pH, 8.27 ± 0.18 and 8.66 ± 0.03, respectively). No correlation was observed between androstenedione RIA and LC–MS/MS discrepancy and lipid or protein. LC–MS/MS sample preparation was tested for the removal of protein-bound material and recovery determined (99–108%). The corresponding RIA results overestimated androstenedione by 52–174% compared to LC–MS/MS. The results here demonstrate that LC–MS/MS is the more accurate method.
Keywords:LC&ndash  MS  Tandem mass spectrometry  Radioimmunoassay  Androstenedione  Stripped serum
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号