Virtual trial to evaluate the robustness of cementless femoral stems to patient and surgical variation
Affiliation:2. Shivaji University, Kolhapur, India;1. Universiti Malaysia Pahang, Faculty of Mechanical Engineering, 26600 Pekan, Pahang, Malaysia;2. Universiti Sains Malaysia, Advanced Medical and Dental Institute, 13200 Bertam, Kepala Batas, Penang, Malaysia;3. Universiti Tenaga Nasional, College of Engineering, Putrajaya Campus, 43000 Kajang, Selangor, Malaysia;4. International Islamic University Malaysia, Kuliyyah of Medicine, 25200 Kuantan, Pahang, Malaysia;5. University of Canterbury, Department of Mechanical Engineering, Private Bag 4800, Christchurch 8041, New Zealand;1. Laboratory of Biomechanics, Justus-Liebig-University Giessen, Klinikstrasse 29, 35392 Giessen, Germany;2. Department of Orthopaedics and Orthopaedic Surgery, University Hospital Giessen and Marburg (UKGM), Klinikstrasse 33, 35392 Giessen, Germany
Abstract:Primary stability is essential for the success of cementless femoral stems. In this study, patient specific finite element (FE) models were used to assess changes in primary stability due to variability in patient anatomy, bone properties and stem alignment for two commonly used cementless femoral stems, Corail® and Summit® (DePuy Synthes, Warsaw, USA). Computed-tomography images of the femur were obtained for 8 males and 8 females. An automated algorithm was used to determine the stem position and size which minimized the endo-cortical space, and then span the plausible surgical envelope of implant positions constrained by the endo-cortical boundary. A total of 1952 models were generated and ran, each with a unique alignment scenario. Peak hip contact and muscle forces for stair climbing were scaled to the donor’s body weight and applied to the model. The primary stability was assessed by comparing the implant micromotion and peri-prosthetic strains to thresholds (150 μm and 7000 µε, respectively) above which fibrous tissue differentiation and bone damage are expected to prevail. Despite the wide range of implant positions included, FE prediction were mostly below the thresholds (medians: Corail®: 20–74 µm and 1150–2884 µε, Summit®: 25–111 µm and 860–3010 µε), but sensitivity of micromotion and interfacial strains varied across femora, with the majority being sensitive (p < 0.0029) to average bone mineral density, cranio-caudal angle, post-implantation anteversion angle and lateral offset of the femur. The results confirm the relationship between implant position and primary stability was highly dependent on the patient and the stem design used.
Keywords:Hip replacement  Implant design  Surgical planning  Finite element modelling  Primary stability  Patient factors  Implant robustness
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号