Quantitation of a new cholecystokinin and gastrin receptor antagonist (L-365,260) in dog and rat plasma by high-performance liquid chromatography |
| |
Authors: | I.-Wu Chen Juinn H. Lin Joan Dorley |
| |
Affiliation: | I.-Wu Chen*, Juinn H. Lin,Joan Dorley |
| |
Abstract: | A high-performance liquid chromatographic (HPLC) procedure has been developed for the quantification of L-365,260 (I), a cholecystokinin and gastrin receptor antagonist, in dog and rat plasma. The method involves liquid—liquid extraction and HPLC with ultraviolet detection. Standard curves were linear over the range 7.5–2000 ng/ml for rat and dog plasma. The method is reproducible and reliable with a detection limit of 7.5 ng/ml in biological fluids. The mean coefficients of variation for concentrations within the range of the standard curve range were 3.84 and 2.56%, respectively, for intra-day analysis and 4.48 and 4.26%, respectively, for inter-day analysis. Application of the development was successfully demonstrated by quantifying the concentration of I in both dog and rat plasma samples following an intravenous or oral dose of 5 mg/kg I. |
| |