首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Pattern ERG: effects of reference electrode site,stimulus mode and check size
Institution:1. Department of Pathology and Laboratory Medicine, University of British Columbia, Vancouver, BC, Canada;2. Department of Pathology and Laboratory Medicine, BC Cancer Agency, Vancouver, BC, Canada;3. Department of Pathology, Royal Columbian Hospital, New Westminster, BC, Canada
Abstract:We studied monocular pattern ERG (PERG) in 10 normal subjects and a patient with optic neuritis. No clinically significant PERG could be recorded from the occluded eye with any reference (ipsilateral ear or temple, or midfrontal), indicating that cross-contamination is not present with binocular testing. Ipsilateral temple reference minimized VEP (P100/N100) contribution to the PERG N95 which occurred with ipsilateral ear or midfrontal reference. The conclusions were confirmed by results from the patient, who had marked monocular delay of a normal amplitude P100. Twenty-four subjects were tested with monocular and binocular stimulation using an ipsilateral temple reference. There were differences in PERG latencies and amplitudes although the interside amplitude ratio showed smaller differences with binocular stimulation. Increasing check size (17, 35 and 70 min) decreased P50 and N95 latencies and increased P50 amplitude.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏