The assay of free potassium in biological buffers with a K+-selective glass electrode
Authors:João PS Cabral
Institution:Center for Experimental Cytology, Oporto University, Rua do Campo Alegre, 823, 4100 Porto, Portugal
Abstract:Abstract

The performance of the Kent K+-selective glass electrode in several biological buffers at neutral pH was evaluated in terms of Nernstian response, repeatability, response time and selectivity. The electrode exhibited a linear response between 2 times 10?5 to 5 times 10?4 and 10?2 M K+, with a slope of 54.9–63.1 mV per decade change in K+ activity. In successive calibrations in the range of 10?5 to 10?2 M K+, the coefficient of variation of the potential in a given K+ concentration decreased with increasing K+ concentration, and was lower than 5%, indicating that in this range of concentrations, the electrode exhibited good repeatability. The response time for a sudden tenfold increase in K+ concentration was 1.3–3.6 min for 10?5 M, and 0.5–1 min for 10?4 M K+. The influence of Ca2+ and Mg2+ on electrode, potential was very small, but Na+ and H+ strongly interfered with electrode response. The selectivity coefficient K+/Na+ was 0.11 and K+/H+ 3.8. The results suggested that in several biological buffers containing no Na+ and with neutral pH, the K+-selective glass electrode can be used to assay with accuracy and rapidity free potassium in the range of 10?5 to 10?2 M, being therefore an alternative to valinomycin-based electrodes.
Keywords:Free K+  K+-selective glass electrode  biological buffers  ion-selective electrode potentiometry