Immobilization of glucose oxidase on thin-film gold electrodes produced by magnetron sputtering and their application in an electrochemical biosensor
Authors:M.H. Gil  J.P. Sardinha  M.T. Vieira  M. Vivan  D. Costa  C. Rodrigues  F.-M. Matysik  A.M. Oliveira Brett
Affiliation:(1) Departamento de Engenharia Química 1, 3000 Coimbra, Portugal;(2) Engenharia Mecânica, 3000 Coimbra, Portugal;(3) Universidade de Coimbra; Hospital da Universidade de Coimbra, 3000 Coimbra, Portugal;(4) Química, 3000 Coimbra, Portugal;(5) Institut für Analytische Chemie, Universität Leipzig, Linnéstrasse 3, D-04103 Leipzig, Germany
Abstract:An electrochemical biosensor is described consisting of a thin-layer gold film electrode prepared by cathodic sputtering using a poly(vinyl chloride) sheet as substrate, with voltammetric behaviour comparable to that of conventional polycrystalline gold electrodes, coated with the hydrolysed copolymer hydroxyethyl methacrylate-co-methyl methacrylate onto which glucose oxidase was immobilized. The mechanical properties of the plastic foil substrate permit easy construction of circular-shaped electrodes which were employed as working electrodes for batch injection analysis. The electrochemical biosensor fabrication is inexpensive and can be used as disposable enzyme sensor for the detection of hydrogen peroxide. The biosensor was tested for the determination of glucose in serum and a good correlation was obtained with the measurement using the electrochemical and the spectrophotometric methods.
Keywords:cathodic sputtering  electrochemical biosensor  glucose  thin-layer gold-sputtered film electrode