High performance liquid chromatography-electrospray ionization mass spectrometric determination of isosorbide 5-mononitrate in human plasma |
| |
Authors: | Sun Xiaohong Li Xiaoqin Cai Shuang Qin Feng Lu Xiumei Li Famei |
| |
Institution: | Department of Analytical Chemistry, Shenyang Pharmaceutical University, Shenyang 110016, PR China. |
| |
Abstract: | A novel, selective and sensitive high performance liquid chromatography-mass spectrometric (HPLC-MS) method has been developed for the determination of isosorbide 5-mononitrate (5-ISMN) in human plasma. With acetaminophen as internal standard, sample pretreatment involved one-step extraction with diethyl ether of 0.5 mL plasma. Analysis was performed on an ACQUITY UPLC BEH C(18) column (100 mm x 2.1mm, 1.7 microm) with mobile phase consisting of acetonitrile-water (20:80, v/v). The detection was carried out by means of electrospray ionization mass spectrometry in negative ion mode with selected ion recording (SIR). Standard curves were linear (r(2)> or =0.99) over the concentration range of 1.04-1040 ng/mL. The lower limit of quantification (LLOQ) was 1.04 ng/mL. The intra- and inter-day precisions (RSDs) were less than 8.6% and 13.4%, respectively, and the accuracy (RE) was within +/-0.45%. The method herein described was fully validated and successfully applied to the pharmacokinetic study of 5-ISMN in compound extended-release tablets in 18 healthy male volunteers after oral administration. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|